Lab Catalog - page 39

sec
SEC
35
Solvent
TSKgel SuperH
6.0 mm ID
×
15 cm
TSKgel H
HR
7.8 mm ID
×
30 cm
n-Hexane
0.5
0.9
methyl ethyl ketone
0.4
0.7
dichloromethane, ethyl acetate
0.35
0.6
toluene, chloroform
0.3
0.5
dimethylformamide
0.2
0.4
carbon tetrachloride, pyridine
0.15
0.3
dimethyl sulfoxide, dioxane, ethanol, N-methylpyrrolidone,
o-dichlorobenzene
0.1
0.2
quinoline, hexafluoroisopropanol, 1-chloronaphthalene
0.05
0.1
table V
Recommended flow rates (mL/min) for TSKgel SuperH and HHR columns
Solvent
Application
THF
polystyrene, epoxy resin, phenoxy resin, polycarbonate, polyisoprene, polyvinyl acetate,
polyvinyl chloride, monoglycerides, fatty acids, polybutadiene, poly(methyl methacrylate),
poly(styrene-butadiene), poly(styrene-acrylonitrile)
n,n-Dimethylformamide (DMF) + 5 mmol/L LiBr
polyvinyl chloride, polyvinyl fluoride, urea resins, polyurethane, polystyrene, polyester, polyimido
ether, polyimido ester, polyphenol (aqueous solution), polyacrylonitrile
o-dichlorobenzene (ODCB)
polyethylene, polypropylene
chloroform
polycarboxylic ether, acrylic resin, epoxy resin, polystyrene
m-cresol/chloroform
nylon, polyester, polyamide, poly (ethylene terephthalate)
toluene
polybutadiene, polysiloxane
table VI
Recommended solvents by application for TSKgel H series columns
Applications of TSKgel H-TYPE Gel PERMEAtion columns
SOLVENTS AND FLOW RATES
TSKgel H series columns can be applied to analyse the molecular mass
distribution of a broad variety of organic-soluble polymers. Table 5 lists
the recommended solvents by application for TSKgel H series columns.
Super H columns are ideally suited to save analysis time and solvent
by semi-micro GPC. For optimum performance they should be used in
combination with a low dead volume GPC instrument such as the all-in-
one EcoSEC system. Table 6 suggests optimum flow rates to be applied
for TSKgel SuperH and TSKgel H
HR
columns for various solvents.