PRODUCTS

TSKgel G5000PW

TSKgel G5000PW

TSKgel G5000PW columns are composed of spherical, hydrophilic polymethacrylate beads. These columns are ideally suited for analysis of water-soluble polymers, such as celluloses, acrylamides, glycols, dextrans, polyvinylalcohol, and oligosaccharides of less than 1 x 10^6 Da molecular weight.

Stable from pH 2 to 12, TSKgel G5000PW columns can be used in mobile phases of water or buffer (up to 20% methanol/80% aqueous) and can tolerate temperatures up to 80 ˚C.


Calibration Curves or TSKgel PW Columns 

PW_Cal-curves.png 
Columns:        TSKgel PW columns, all 7.5 mm ID × 60 cm
                         A. G2000PW, B. G2500PW, C. G3000PW, D. G4000PW, 
                         E. G5000PW, F. G6000PW, G. GMPW
Mobile phase: distilled H20
Flow rate:       1.0 mL/min
Detection:        RI 

 

Ordering Information

P/N Description Dimension
0005764 TSKgel G5000PW 7.5 mm ID x 30 cm L
0005108 TSKgel G5000PW 7.5 mm ID x 60 cm L
0006762 TSKgel PWH Guardcolumn 7.5 mm ID x 7.5 cm L