TSKgel G5000PWXL-CP
The TSKgel G5000PWXL-CP column was specifically developed for the analysis of water-soluble cationic polymers with molecular weights up to 1 x 10^6 Da. Composed of 10 µm polymethacrylate beads, cationic groups are introduced on the surface of the TSKgel G5000PWXL-CP packing material to prevent adsorption of cationic polymers and allow elution under low salt conditions.
This columns shows high theoretical plate numbers, linear calibration curves and high durability because the base resin is the same as that used in the TSKgel PWXL series columns.
Polyethylene glycol and oxide calibration curves for TSKgel PWXL-CP columns |
|
Columns: TSKgel G3000PWXL-CP, 7 µm, 7.8 mm ID × 30 cm TSKgel G5000PWXL-CP, 10 µm, 7.8 mm ID × 30 cm TSKgel G6000PWXL-CP, 13 µm, 7.8 mm ID × 30 cm Mobile phase: H2O with 0.1 mol/L NaNO3 Flow Rate: 1 mL/min Detection: RI Temperature: 25 °C Samples: polyethylene oxides (PEO) standards polyethylene glycols (PEG) standards |
Analysis of cationic polymers |
|
Columns: TSKgel G3000PWXL-CP, 7 µm, 7.8 mm ID × 30 cm TSKgel G5000PWXL-CP, 10 µm, 7.8 mm ID × 30 cm TSKgel G6000PWXL-CP, 13 µm, 7.8 mm ID × 30 cm Mobile phase: H2O with 0.1 mol/L NaNO3 Flow Rate: 1 mL/min Detection: RI Temperature: 25 °C Sample Load: 3 g/L, 100 µL |
Analysis of PAA |
|
Column: TSKgel G5000PWXL-CP, 10 µm, 7.8 mm ID × 30 cm Mobile phase: H2O with 0.1 mol/L NaNO3 Flow rate: 1.0 mL/min Detection: RI Temperature: 25 °C Samples: polyallylamine-HCl (PAA) Sample load: 3 g/L, 100 µL |