TSKgel G2000SW
As the first generation of silica-based columns for aqueous SEC, TSKgel G2000SW columns provide excellent separation of peptides and proteins with molecular weights up to 100,000 Da.
TSKgel G2000SW columns are available in 10 and 13 μm for analytical and semi-preparative purifications.
Separation of nucleic acid
|
|
Column: TSKgel G2000SW, 10 µm, 7.5 mm ID × 60 cm
Mobile phase: acetic acid/triethylamine/H2O = 3/3/94
Flow rate: 0.74 mL/min
Detection: UV @ 260 nm
Samples: 1. uridine, 2. uracil
3. thymine, 4. adenosine
5. adenine |