PRODUCTS

TSKgel G5000PWXL-CP

TSKgel G5000PWXL-CP

The TSKgel G5000PWXL-CP column was specifically developed for the analysis of water-soluble cationic polymers with molecular weights up to 1 x 10^6 Da. Composed of 10 µm polymethacrylate beads, cationic groups are introduced on the surface of the TSKgel G5000PWXL-CP packing material to prevent adsorption of cationic polymers and allow elution under low salt conditions.
This columns shows high theoretical plate numbers, linear calibration curves and high durability because the base resin is the same as that used in the TSKgel PWXL series columns.

Polyethylene glycol and oxide calibration curves for TSKgel PWXL-CP columns 

PWXL-CP_fig1.png 
Columns:        TSKgel G3000PWXL-CP, 7 µm, 7.8 mm ID × 30 cm
                         TSKgel G5000PWXL-CP, 10 µm, 7.8 mm ID × 30 cm
                         TSKgel G6000PWXL-CP, 13 µm, 7.8 mm ID × 30 cm
Mobile phase: H2O with 0.1 mol/L NaNO3 
Flow Rate:      1 mL/min
Detection:        RI
Temperature:  25 °C
Samples:         polyethylene oxides (PEO) standards
                         polyethylene glycols (PEG) standards 


Analysis of cationic polymers 

PWXL-CP_fig2.png 
Columns:        TSKgel G3000PWXL-CP, 7 µm, 7.8 mm ID × 30 cm
                         TSKgel G5000PWXL-CP, 10 µm, 7.8 mm ID × 30 cm
                         TSKgel G6000PWXL-CP, 13 µm, 7.8 mm ID × 30 cm
Mobile phase: H2O with 0.1 mol/L NaNO3 
Flow Rate:       1 mL/min
Detection:        RI
Temperature:   25 °C
Sample Load:  3 g/L, 100 µL 


Analysis of PAA 

PWXL-CP5000_fig3.png
 
Column:           TSKgel G5000PWXL-CP, 10 µm, 7.8 mm ID × 30 cm
Mobile phase:  H2O with 0.1 mol/L NaNO3
Flow rate:        1.0 mL/min
Detection:        RI
Temperature:  25 °C
Samples:          polyallylamine-HCl (PAA)
Sample load:   3 g/L, 100 µL

 

Ordering Information

P/N Description Dimension
0021874 TSKgel G5000PWXL-CP 7.8 mm ID x 30 cm L
0021876 TSKgel PWXL-CP Guardcolumn 6.0 mm ID x 4.0 cm L