The TSKgel SuperH2500 GPC column is packed with rigid, porous, polymer beads. It is for the non-aqueous separation of oligomers and polymers with molar masses of less than 20,000. The maximum operating temperature for the TSKgel SuperH2500 column is 140 °C. This column is shipped in THF, which can be exchanged for a wide variety of organic solvents.

TSKgel SuperH2500 is a high efficiency/high throughput version of the conventional TSKgel G2500HHR column. Both column types are based on the same bead chemistry. The volume of the 6 mm ID × 15 cm TSKgel SuperH2500 column is 3.4 times smaller than that of a conventional 7.8 mm ID × 30 cm column. As a result, peak volumes will be proportionally smaller on this column compared to a corresponding TSKgel HHR column. Thus, your HPLC system may require optimization of components that can give rise to extra-column band broadening, such as connecting tubing, injector, injection volume, detector cell volume, and detector time constant.

Separation of standard polystyrenes using a TSKgel SuperH2500 column 

superh2500_fig1.png 
Column:          TSKgel SuperH2500, 3 µm, 6 mm ID × 15 cm
Mobile phase: THF, chloroform, DMF, carbon tetrachloride 
Flow rate:        0.6 mL/min 
Temperature:  25 °C 
Detection:        UV/VIS @ 254 nm or 270 nm  
Samples:          1) polystyrene (1.9 × 10^5 Da), 2) polystyrene (9.1 x 10^4 Da)
                          3) polystyrene (2,800 Da), 4) polystyrene A-500