TSKgel SuperH3000

The TSKgel SuperH3000 GPC column is packed with rigid, porous, polymer beads. It is for the non-aqueous separation of polymers with molar masses of less than 60,000. The maximum operating temperature for the TSKgel SuperH3000 column is 140 °C. This column is shipped in THF, which can be exchanged for a wide variety of organic solvents.

This TSKgel SuperH3000 column is a high efficiency/high throughput version of the conventional TSKgel G3000HHR column. Both column types are based on the same bead chemistry. The volume of the 6 mm ID × 15 cm TSKgel SuperH3000 column is 3.4 times smaller than that of a conventional 7.8 mm ID × 30 cm column. As a result, peak volumes will be proportionally smaller on this column compared to a corresponding TSKgel HHR column. Thus, your HPLC system may require optimization of components that can give rise to extra-column band broadening, such as connecting tubing, injector, injection volume, detector cell volume, and detector time constant.

Temperature dependence of separation on epoxy resin 

superh3000_fig1.png 
Columns:         TSKgel SuperH3000, 3 µm, 6 mm ID × 15 cm × 2 +
                          TSKgel SuperH2500, 3 µm, 6 mm ID × 15 cm × 3
Mobile phase: THF
Flow rate:        0.6 mL/min
Detection:        UV @ 254 nm
Sample:           Epikote 1004 (0.1%), 10 µL